Tinzaparin

LMWH, Anti XA, Low Molecular Weight Heparin
Core Laboratory & Transfusion Medicine
Hematology
Blood
TINZA

MSH Coagulation requisition form (MS 520)

This chromogenic assay is performed for the quantitative determination of the Low Molecular Weight Heparin (LMWH) level.

Plasma (3.2% Na Citrate)

Full draw

Light Blue citrate tube
Collection Containers – Tube Types

The 9:1 blood/citrate ratio must be kept constant. Adjustments for the amount of anticoagulant should be made when the hematocrit is abnormal (above 55%). Mix immediately to prevent clotting.
Collection Instructions

Send whole blood sample(s) at room temperature as soon as possible.
If specimens cannot be shipped immediately, centrifuge and separate plasma within 1 hour of collection, send frozen plasma on dry ice

Fridge: N/A
Centrifuge and separate plasma within 1 hour of collection.
Room Temperature: 4 hours
Freezer: - 20 °C 15 days - 70 °C 6 months


Automated Coagulation Analyzer

24 hours, 7 days/week.

2 hours

4 hours

Suggested LMWH Therapeutic Ranges:
1.0-2.0 IU/mL once daily dosage
0.6-1.0 IU/mL twice daily dosage In case of discrepancies, always interpret results using reference ranges as stated on laboratory reports and/or electronic patient chart.

416-586-4800 Extension 4688

Must indicate anticoagulant status